Model AAS-1E2 / 1E3 Atomic Absorption Spectrophotometer

FEATURES

           Perfect optical system with high energy.

           Various atomization systems including flame, furnace and hydride are available for extending analytical functions Innovative design of the electronics microcomputer system  with high stability and reliability.

           Complete functions and easy operation.

           Background correction adopts the advanced deuterium arc background correction system with strong correction capability.

           Various kinds of accessories and spare parts are available for enhancing use of the instrument.

 

 

aas-1E2 aas-1E2

MAIN SPECIFICATIONS:

 

     Wavelength range:190-900nm

     Wavelength accuracy:\0.5nm

     Resolution: Better than 0.3nm

     Baseline Stability: \0.005A/30min

     Background absorption up to 1A can be corrected.

TECHNICAL SPECIFICATIONS

Light Source System

           Primary and auxiliary source is modulated by 400HZ pulse with time-shared system.

           Lamp current is set by the microcomputer. Duty cycle of supplied current is 1:5.

           The mean current is 0.0-20mA (max. peak current can be reached up to 200mA).

           The deuterium lamp current can be auto-adjusted. Current range is from 70 to 100mA.

           The two hollow cathode lamps can be lit simultaneously.

Optical System

           Photometric system: Single-beam, dual-lens optics.

           Monochromator Ebert design, grating monochromator Plane diffraction grating with 1200L/mm, blazed at 250nm, spectral bandwidth: 0.2,0.4,1.0,2.0nm.

Flame Atomization System

           5cm and 10cm changeable all-titanium, single-slot burner head.

           The chamber is internally coated with corrosion-resistant plastic.

           Provide adjustable metal nebulizer and high-efficiency glass nebulizer with metal sheath.

           Gas control system uses pressure gauge, flow meters with needle valve, pressure-stabilizer for monitoring and adjusting both fuel and oxidant gases and has an air-N2O changeover and flashback preventing device.

           Adjusting mechanism of the vertical, horizontal positions and rotating angle for burner head.

           Auto-igniting mechanism. (optional)

Detection System

Photomultiplier with high sensitivity and wide spectral range is used as signal detector.

Electronic-Microcomputer System

         The source energy and high voltage of photomultiplier can be automatically adjusted and balanced.

           Energy and measuring values are displayed on LED.

           The signals of atomic absorption, background correction or background can be detected.

           Signal readout mode includes transient, integral, peak height, peak area.

           Selected integral or sampling time is in the range of 0.1-19.9s.

           Reading delay time can be selected from 0 to 9.9 sec.

           Scale expansion is from 0.1x to 99x.

           All readings and statistics including mean, standard deviation and relative standard deviation can be automatic calculated.

           Selected repeating number is in the rang of 1-99.

           Working curve is automatic fitted with three standard.

           The concentration of sample can be directly read.

           The slope of working curve with a standard can be re-calibrated.

           Printing functions: totally or partially printout measurement data, calibration curve, graphics and analytical conditions.

           The parameters set up by each function key can be checked.

           Built-in special parallel interface, compatible computer can be connected.

ENVIRONMENTAL REQUIREMENTS

           Operating Temperature: 5-35<C

           Relative Humidity: Less than 85%

           Power Supply: 220V/110V, 50/60Hz, 220W,Single phase

           Dimension: 1150(L)〜370(D)〜 510(H)mm

           Weight: 110Kg

 

More functions of model AAS-1E3: Fast scanning of wave length, auto protection of gas channel

 

Model AAS-120 / 130 Atomic Absorption Spectrophotometer

FEATURES

SPECIFICATIONS

 

ENVIRONMENT REQUIREMENT

Ambient temperature: 10-30
Humidity: less than 85%
Power requirement: 220v, 50Hz, 5A

AAS-100

DIMENSION AND WEIGHT

Dimension: 960 (L) X 510 (D) X 550 (H) mm
Weight: unpacked 70kg

WF-1C GRAPHITE FURNACE POWER SUPPLY

Features

Technical Specifications

SYSTEM CONTROL AND DATA PROCESSING

Main functions of AAS-130: 
auto switch of 4-lamp turret, auto set of wave length, auto switch of
Spectral bandwidth,
Auto-igniting mechanism, total automatic gas protection, and deuterium arc background correction system.
  Others are same as AAS-120