FEATURES

- Digital filter, free of AC, drift and EMG interference

- High resolution thermal printer head

- Display ECG trace and operation menu in large LCD screen, set parameters freely

- Simultaneous acquisition ECG parameters. Simultaneous printing (ECG-1212 only)

- Automatic Baseline Control

- AC/DC, DC can use for 50 patients

- Soft one- tough operation

- Selectable 3-channel, 3-channel+any rhythm lead, 6-channel, 12-channel (ECG-1212 only)

- Standard input/output, and RS232 interface

- Automatic measurement and analysis of ECG parameters.(optional)

- Records management system is optional.

- Data and analysis results storable (for ECG-1206 is optional)
- Paperless data collection, analysis, re-play and copy (optional)

SPECIFICATIONS

Input Circuit

Isolated

Sensitivity

5mm/mV,10mm/mV,

Safety Level

Class I,Typa CF

20mm/mV,Auto

Input Impedance

50MΩ

Display

320x240 chart LCD display

Input Circuit Current

50nA

Paper Speed

25mm/s,50mm/s

Noise Level

15μVP-P

Paper size

Rolling type 112mmx30m or folded type 112mmx75mm (ECG-1206) 210mmx30m or 210mmx140mmx20m ( ECG-1212)

Patient Current Leakage

10μA

Power Requirements

AC:220V,50Hz,110V,60Hz

CMRR

100dB

DC: Rechargeable

Frequency Response

0.05-150Hz(-3dB)

 

battery supporting 50 exams

A/D Constant

12 bit

Weight

<3.5Kg (including battery) ECG-1206

Time Constant

>3.2 second

<5Kg (including battery) ECG-1212

Sampling Rate

1ms

Dimension

327X252X76mm (ECG-1206)

Despolarization Voltage

±300mV

400x325x84.5mm (ECG-1212)

Standard Accessory

 

 

 